Rod and Schwartz spectrometer FSV - k30 connected via LAN remote control

Time:09-17

So connected to a port number and the spectrometer collection software only IP Settings, no port Settings, so the front-end ports to connect how connection??? Spectrometer how to set the port number???

CodePudding user response:

That would be to use fixed port, look at its manual,

CodePudding user response:

Manual is finished, there is no port!!! I want to say this instrument to open a XP system, built-in data acquisition software, is equal to no port connection words don't know is for the data acquisition software to send, so send the manual remote control instruction, the acquisition software have no reaction,,,,

CodePudding user response:

If not found, it is self agreement, IP address of the can, had better look for initializing...

CodePudding user response:

Connection is equal to connect with computer, the port number I use 135 remote control port, is not to know to send data, which sent a instruments have no reaction, the what, who,, to say ah,,,,,,,,,,

CodePudding user response:

refer to the second floor Fitable_miss response:
manual finished, there is no port!!!!!! I want to say this instrument to open a XP system, built-in data acquisition software, is equal to no port connection words don't know is for the data acquisition software to send, so send the manual remote control instruction, the acquisition software have no reaction,,,,


If the instrument can install the software, use tcpview or Windows built-in netstat see it time and software to monitor the which port, if you can't install software, network port scanning software in other computer after such as nmap to sweep the device can be connected port,

CodePudding user response:

First of all, installation aglient IO suite;
Then, find the FSV - K30 programming manual, according to the programming manual operation;
With so connection is don't need the port number, its IP connection parameter is the format 'TCPIP: : 192.168.0.1
'Using GPIB connection, the parameter format is' GPIB: : 28 ',
Now see a programming manual, all instrument connections, are NI a set of standards,

CodePudding user response:

Eldest brother, connecting the spectrometer under the specific steps to say, thank you. I now is: 1, open the spectrum instrument, spectrometer XP system automatically open the acquisition data of the software; 2, the inside of the acquisition software to change the IP is the same and my computer XP system to change the IP, it's easy to get the IP address and change; 3, open the computer, open the serial debugging software, can instrumentation, general network connections are sweet, sweet, slogans, (don't know the port number, so I'll use port 135 remote control, can be connected) 4, send instruction to the instrument with serial debugging software (I sent directly is the change of frequency & lt; FREQ: STAR 20 MHZ & gt; , change the bandwidth & lt; INP: ATT 30 db & gt; To return to the instrument information instruction & lt; The * IDN?>; The instrument have no reaction, what is connected to the instrument to start after a command, the can send other control instructions)

CodePudding user response:

How can't port, should be fixed port, 1000, for example, you use 135 must not, all of these, 80, and the port like using the assigned to a specific type of application,